MAU01M Series
1W DC-DC Converters
Features
  • Industrial Standard SIP-7 Package
  • Unregulated Output Voltage
  • I/O Isolation 4000VAC with Reinforced Insulation, rated for 300Vrms Working Voltage
  • Low Leakage Current
  • Operating Ambient Temp. Range -40 ℃ to 95 ℃
  • Short Circuit Protection
  • Medical EMC Standard with 4th Edition of EMI EN 55011 and EMS EN 60601-1-2 Approved
  • Medical Safety with 2xMOPP per 3.2 Edition of IEC/EN 60601-1 & ANSI/AAMI ES60601-1 Approved with CE Marking
OVERVIEW

Introducing the MINMAX MAU01M series – an innovative range of 1W medical-approved isolated DC-DC converters encapsulated in a SIP-7 package, meticulously designed for medical applications. With 9 models available, supporting input voltages of 5, 12, and 24VDC, and providing output voltages of 5, 12, and 15VDC, this series ensures versatility to meet various medical device requirements.
The MAU01M series boasts an I/O isolation specified for 4000VAC with reinforced insulation, rated for a reliable 300Vrms working voltage. Additional features include short circuit protection, low I/O leakage current of 2μA max, and an operating ambient temperature range from -40℃ to 95℃ without derating. Aligning with the 4th edition medical EMC standard, the series holds medical safety approval with 2xMOPP (Means Of Patient Protection) per the 3.2 Edition of IEC/EN 60601-1 & ANSI/AAMI ES 60601-1.
In adherence to ISO 14971 Medical Device Risk Management, the MAU01M series undergoes a comprehensive risk assessment process. This ensures not only compliance with high-performance standards but also alignment with the rigorous safety benchmarks outlined in ISO 14971. By seamlessly integrating the MAU01M series into medical devices, you not only benefit from its compact design and versatile voltage options but also ensure compliance with comprehensive risk management protocols.
In summary, the MAU01M series offers an optimal solution for demanding applications in medical instruments, now fortified with the assurance of ISO 14971 compliance. Elevate your medical devices with the MINMAX MAU01M series – where innovation meets safety, performance, and meticulous Medical Device Risk Management Report Acquisition.

Specifications
Line Regulation ±1.2% typ.
Load Regulation
(Io = 10% - 100% load change)
±10% max.
Ripple & Noise (20MHz) 75mVp-p max.
Efficiency Up to 81%
I/O Isolation Voltage 4000VAC min. (Reinforced Insulation)
I/O Isolation Capacitance 20pF typ.
I/O Leakage Current 2µA max.
Operating Ambient Temp. Range -40℃ to +95℃ (See Derating Curve)
Short Circuit Protection Continuous (Auto. Recovery)
Case Material Plastic (UL94V-0 rated)
Safety Approval 2xMOPP per 3rd Edition of IEC/EN 60601-1 & ANSI/AAMI ES60601-1,CB-Report, CE Marking
Contact local distributor for quote

Comparison ECAD Model Number Output Power Input Voltage (VDC) Output Voltage (VDC) Output Current (mA) Efficiency Output Regulation I/O Isolation Package Type