MIW03M Series
3.5W DC-DC Converters
Features
  • Industrial Standard DIP-24 Package
  • Wide 2:1 Input Voltage Range
  • Fully Regulated Output Voltage
  • I/O Isolation 5000VAC with Reinforced Insulation, rated for 250Vrms Working Voltage
  • Creepage & Clearance Distance meet 8mm
  • Low Leakage Current
  • Operating Ambient Temp. Range -40 ℃ to 96 ℃
  • No Min. Load Requirement
  • Under-Voltage, Overload/Voltage and Short Circuit Protection
  • Conducted EMI EN 55011 Class A Approved
  • Medical EMC Standard with 4th Edition of EMI EN 55011 and EMS EN 60601-1-2 Approved
  • Medical Safety with 2xMOPP per 3.2 Edition of IEC/EN 60601-1 & ANSI/AAMI ES60601-1 Approved with CE Marking
OVERVIEW

Introducing the MINMAX MIW03M series – an innovative range of high-performance 3.5W medical-approved DC-DC converters encapsulated in a DIP-24 package, purposefully designed for medical applications. With an extensive selection of 21 models supporting input voltages of 5, 12, 24, and 48VDC, featuring a wide 2:1 input range and fixed output voltage, this series ensures adaptability to diverse medical device specifications.
The MIW03M series boasts an I/O isolation specified for 5000VAC with reinforced insulation, rated for a reliable 250Vrms working voltage. Advanced features include under-voltage, overload, over-voltage, and short-circuit protection, along with no minimum load requirement, conducted EMI EN 55011 class A approval, low I/O leakage current of 2µA max, and an operating ambient temperature range from -40℃ to +96℃ without derating, achieved through high efficiency up to 87%.
Aligned with the 4th edition medical EMC standard, the MIW03M series holds medical safety approval with 2xMOPP (Means Of Patient Protection) per the 3.2 Edition of IEC/EN 60601-1 & ANSI/AAMI ES 60601-1, incorporating an 8mm creepage and clearance.
In adherence to ISO 14971 Medical Device Risk Management, the MIW03M series undergoes a thorough risk assessment process. This ensures not only compliance with high-performance standards but also alignment with the stringent safety benchmarks outlined in ISO 14971. Elevate your medical devices with the MINMAX MIW03M series – the integration of advanced technology, safety, performance, and Medical Device Risk Management Report Acquisition.

Specifications
Output Voltage Accuracy ±1.0% max.
Line Regulation ±0.5% max.
Load Regulation ±0.5% max.
Ripple & Noise (20MHz) 70mVp-p typ.
Efficiency Up to 87%
I/O Isolation Voltage 5000VAC min. (Reinforced Insulation)
I/O Isolation Capacitance 20pF max.
Operating Ambient Temp. Range -40℃ to +96℃ (See Derating Curve)
Short Circuit Protection Continuous (Auto. Recovery)
Case Material Plastic(UL94V-0 rated)
Safety Approval  2xMOPP per 3rd Edition of IEC/EN 60601-1 & ANSI/AAMI ES60601-1,CB-Report , CE Marking
Contact local distributor for quote

Comparison ECAD Model Number Output Power Input Voltage (VDC) Output Voltage (VDC) Output Current (mA) Efficiency Output Regulation I/O Isolation Package Type